Program Structure; Adjustable Parameters And Possible Value Ranges - Ivoclar Vivadent Programat P300 Operating Instructions Manual

State-of-the-art furnace for dental applications
Table of Contents

Advertisement

5.3 Program structure

The furnace offers three types of programs:
a. Standard programs for Ivoclar Vivadent
materials
b. Free programs
c.
Auxiliary programs
All individual programs are equivalent and therefo-
re full-fledged programs. In each program, all the
parameters can be adjusted.
a) Standard programs for Ivoclar Vivadent
materials (see chapter 10.1)
– IPS d.SIGN
– IPS InLine
– IPS Empress Esthetic
– IPS e.max Ceram
– Various Add-On materials
b) Free programs
– Free, individually adjustable programs
When the furnace is delivered ex
works, the standard programs already
contain the recommended material
parameter settings.
However, the parameters can be changed and
overwritten at any time, if required, if the
programs are to be used for other purposes.
Therefore, the 50 programs are also available as
free programs.
The programs are designed in such a way that
they can be either used as conventional, one-stage
programs or as two-stage programs, if required.
The mode can be changed via the symbol (one- or
two-stage program) by using the + or – key.
c.
Auxiliary programs
Various test programs are available. Please refer
to chapter 5.5 Settings / configuration and
information.

5.4 Adjustable parameters and possible value ranges

Symbol
Parameter
P
Program number P
B
Stand-by temperature
S
Closing time (min : sec)
Temperature increase rate
T
Holding temperature
H
Holding time (min : sec)
V1
Vacuum on
V2
Vacuum off
Temperature increase rate
t2À
Second stage
T
Holding temperature
Second stage
H
Holding time
Second stage (min : sec)
V1 (V1 2)
Vacuum on
Second stage
V2 (V2 2)
Vacuum off
Second stage
L
Long-term cooling
t L
Cooling temperature rate
**) The maximum value of the cooling temperature rate is adjusted to the set holding temperature T and long-term
cooling L (e.g. with T 1180, L 1000 -> max. t
Automatic plausibility check
The furnace is equipped with an automatic plausibility check function. The parameters
(e.g. T 960 but L 1000) are checked upon each program start. In case of contradictory
parameter combinations, the program stops automatically and the respective error
number is indicated.
19
Value range
Value range
1–100
100–700 °C
212–1292 °F
00:18–30:00
30–140 °C/min
54–252 °F/min
100–1200 °C
212–2192 °F
00.01–60:00
0 or 1–1200 °C
0 or 34–2192 °F
0 or 1–1200 °C
0 or 34–2192 °F
30–140 °C/min
54–252 °F/min
100–1200 °C
212–2192 °F
00.01–60:00
0 or 1–1200 °C
0 or 34–2192 °F
0 or 1–1200 °C
0 or 34–2192 °F
0 or 50–1200 °C
0 or 122–2192 °F
0 or 1–50 **)
0 or 32–90
À
100 °C possible, but with T 850 L 700 -> max. t
À
50 °C)